Correlating Your Lab With the Supplier Lab: Making Two Correct Measurements Agree

Author: CDTech Publish Time:  Origin:

Two labs measure the same panel and report numbers a few percent apart. Both are calibrated, both follow a method, and the argument that follows is usually about trust rather than physics. The gap is normally a difference in what was measured, not in how well. This article sets out the conditions that must match before a comparison means anything, the correlation study that turns a disagreement into a documented offset, and the cases where the honest answer is to accept the other lab's number.

Why two correct measurements disagree

Because the setups measure different quantities.

Calibration makes an instrument right, not identical. Aperture, distance, warm-up, field size and the point measured on a non-uniform panel change the number before any error exists.

A luminance reading is not a property of the panel alone. It is the output of a system: an instrument with an aperture, a distance, a focus, a warm-up state and a signal chain, aimed at a point on a surface that is not perfectly uniform. Change the aperture from a narrow spot to a wide one and the meter averages a different area. Change the distance and the collected cone changes. Measure before the backlight has stabilised and the number is still climbing.

Two labs can therefore be individually correct and jointly inconsistent. Neither is wrong about its own measurement; they are wrong to treat the two numbers as the same quantity. The first step in any dispute is to establish whether the setups measured the same thing, and that is a documentation question before it is a metrology question.

The measurement conditions that must match before any comparison

Before comparing numbers, compare methods line by line. The parameters below change a luminance or chromaticity result even when both instruments are calibrated and both operators are competent. Most correlation failures are found in this list rather than in the instrument.

ParameterWhat to fixEffect if mismatched
Warm-up and driveStated warm-up at rated currentCold readings run low
Ambient temperatureRecorded and matchedOutput and colour drift with heat
Aperture and fieldSame aperture angleA different area is averaged
Distance and focusSame distance, focusedThe collected cone changes
Viewing angleNormal to the panelOff-axis readings fall away
Measurement pointsMarked on the sampleNon-uniform panel, different area
Signal and patternSame level, gamma, patternBrightness and grey level change
Instrument typePhotometer or spectroradiometerFilter mismatch on narrow spectra

Two further items belong on the comparison sheet: the calibration certificate number of each instrument, and the software version used to reduce the data. Test methods for the optical quantities themselves come from bodies such as ASTM.

Instrument, aperture and distance: the three biggest levers

The instrument type decides what is being averaged. A filtered photometer matches the eye closely but drifts with filter age, while a spectroradiometer measures the spectrum and computes luminance from it. On a narrow spectrum, a filter that is slightly off produces a bias calibration alone will not remove, and the display measurement standards in the IEC webstore define such readings.

Aperture is the lever most often overlooked. A wide spot and a narrow spot average different areas, and on a panel with a bright centre or a dim corner the two can differ by more than either instrument's error. The aperture belongs in the method, not the operator's judgement.

Distance and focus set the cone the instrument collects. Moving a meter back changes the solid angle and the area sampled, and a defocused lens collects light from a wider region. The optics and backlight guide covers how the panel's uniformity interacts with those choices.

10.1 inch IPS TFT LCD display with CTP and optical bonding, 600 nits
A 10.1 inch IPS TFT LCD display with CTP and optical bonding, 600 nits. Optical bonding removes an air gap, so luminance and contrast read higher than on the same panel without it.

Running a correlation study on the same sample set

A correlation study is a measurement exercise with a fixed protocol, not a meeting where numbers are compared. The protocol below is the minimum that produces a result both labs can defend.

  • Define the measurand in words: luminance at the centre, at a stated current, after a stated warm-up.
  • Use one sample set of units covering the range, not a single convenient panel.
  • Mark the measurement points on each unit so both labs read the same place.
  • Fix the drive and pattern with a written signal setup, not a remembered one.
  • Log the environment: ambient temperature, warm-up time, instrument and aperture.
  • Record raw values with instrument ID and calibration certificate number.
  • Exchange data before discussion, so neither lab adjusts to the other's numbers.
  • Repeat each reading at least twice to separate repeatability from disagreement.

Five or more units spanning the working range is a common starting point; the number rises with the scatter and with how far the result must hold. A study built on one panel proves only that two instruments can read one area on one day.

Reading a disagreement: bias, slope or scatter

When the data are plotted, the shape of the difference matters more than its size. A constant offset across the range is a bias: the two setups differ by a fixed amount, and a documented offset can absorb it. A difference that grows with brightness is a slope: one setup reads proportionally higher, which usually points at aperture, gain or geometry rather than at drift.

Scatter is the third pattern and the one that cannot be corrected by arithmetic. If the points fall either side of a line with a wide spread, the two labs are not measuring the same thing consistently, and an offset will simply hide the problem. Look at repeatability first: two readings from one lab on one unit should agree closely before any cross-lab comparison is attempted.

The NIST guidance on measurement uncertainty is useful here, because it forces the question of what the combined uncertainty of the comparison actually is. A gap smaller than that uncertainty is not evidence of a fault.

Gauge R and R for a visual check and an optical check

Gauge R and R answers a narrower question than correlation: how much of the variation comes from the measurement system rather than the parts. For an optical check it is a variables study, with instrument, operator and sample position as sources. For a cosmetic check it is an attribute study, with inspectors grading the same panels twice against the same reference.

In both cases the study needs parts that span the tolerance, not a set of identical good units. A study on ten perfect panels produces a flattering number that predicts nothing, because there is no variation for the measurement system to obscure. The convention that a gauge study should contribute under a tenth of the tolerance, and is questionable above about a third, comes from the automotive measurement systems literature published by AIAG.

Optical studies carry a specific trap: reproducibility across operators often reflects a different bench, a different warm-up or a different aperture rather than a different pair of hands. Record the setup with each operator's data, or the study will blame people for equipment.

Writing the correlation result into the acceptance clause

A correlation that lives in an email dies at the next audit. The result belongs in the acceptance clause as a documented relationship: the measurand, the method and conditions, the instrument class, the measurement points, and the offset or correction that applies between the two labs.

State the range over which the offset holds. An offset derived between two brightness levels should not be applied at a third without evidence, and a correction that works on a bonded module may not transfer to an unbonded one. Say what happens outside the range: the measurement is referred back to the lab that owns the method rather than corrected by extrapolation.

The clause should also name the reference lab and the reason for that choice, which is usually the method's origin rather than commercial convenience, and note that the arrangement sits inside a documented quality system such as ISO 9001.

Re-correlating after an instrument or fixture change

A correlation is a snapshot of two setups, and it expires when either setup changes. The triggers should be written with the clause: a new or repaired instrument, a change of aperture, a new fixture or sample holder, a software or firmware update that alters the reduction, a move to another room with different lighting or temperature control, and a change of operator where only one person was trained.

Recalibration is not the same event. A returning instrument carries a fresh certificate but may still differ from the setup behind the original correlation, particularly if the repair touched the optical path. Re-run the study on the retained sample set rather than assuming the certificate restores the relationship.

10.1 inch PCAP and optical bonding TFT display with anti-glare treatment
A 10.1 inch PCAP and optical bonding TFT display with anti-glare treatment. An anti-glare surface scatters light, so reflectance and contrast readings depend on the geometry both labs use.

Keeping a set of correlation panels is what makes re-running cheap. Mark them, store them away from light and heat, and treat them as reference artefacts rather than saleable stock. The wider record discipline is covered in the quality and reliability guide.

When to accept the supplier number instead of arguing

When the gap sits inside combined uncertainty.

If the difference is smaller than what the two labs' uncertainty budgets allow, neither number is wrong. Spend the effort on the process rather than on the argument.

There are also cases where the supplier's number wins on method rather than on statistics. If the drawing, the customer specification or the purchase order names a method the supplier follows and your lab approximates, the supplier's reading is the conforming one, and yours should be reported as an indication. The same applies when the sample sits at the edge of your instrument's range, or when the correlation never covered the brightness level in dispute.

Accepting a number is not the same as accepting a defect. The two decisions are separate: agree the measurement, then argue about the specification. Conflating them turns a two percent difference into a stopped shipment, and the calibration scope behind each lab's certificate is set out on the quality certifications page. Incoming inspection sampling, as distinct from measurement agreement, is covered in the incoming inspection guide.

Which plot should two laboratories use to show they agree?

Bland-Altman, not a correlation coefficient.

A high correlation between two labs proves they move together, not that they agree. Two instruments can correlate at 0.99 while one reads eight per cent higher throughout, and that constant offset is precisely the error an acceptance limit will catch. The plot that answers the real question is the difference against the mean.

Plot each sample with the mean of the two readings on one axis and their difference on the other. The mean difference is the bias; the spread gives the 95 per cent limits of agreement. If those limits sit inside your acceptance band, the labs are interchangeable for that characteristic. If they do not, no correlation figure rescues the comparison, and the discussion moves to which laboratory the specification names.

If your team is correlating a lab against a supplier's and needs the measurement conditions, raw data and instrument records behind the numbers, CDTech can supply those with the samples and the calibration documentation. Send the method you are working to and the values in dispute to [email protected], or use the contact page to set out what the comparison needs to cover.

FAQ

Why do two luminance meters disagree?

Usually because they are measuring different areas under different conditions rather than because one is broken. Aperture, distance, focus, warm-up and the point chosen on a non-uniform panel all change the reading, and a filtered photometer can also differ from a spectroradiometer on a narrow spectrum. Check the method line by line before suspecting the instrument, then run a correlation study on a shared sample set so the difference can be described as bias, slope or scatter.

How many samples does a correlation need?

Enough to cover the range over which the result will be used, which in practice means units at the low, middle and high end of normal output rather than several identical good panels. Five units is a common minimum, and the count should rise with the scatter and with the consequences of being wrong. Repeat readings on each unit as well, so repeatability can be separated from the disagreement between labs.

Does correlation replace calibration?

No. Calibration establishes that an instrument reads correctly against a traceable reference; correlation establishes how two specific setups relate to each other. A calibrated instrument can still disagree with another calibrated instrument because the methods differ. Keep both: certificates for the instruments, a correlation record for the pair of labs, and a re-correlation trigger for any change that could move either setup.

Can correlation be done remotely?

Partly. A shared sample set can travel between labs, which supports a study without a visit, provided the units are robust, marked and measured at identical points under logged conditions. What travels poorly is the setup: aperture, distance and fixturing are easier to verify in person or by video walkthrough than by description. A remote correlation is workable if the protocol is explicit and both labs exchange raw data and photographs of the bench before comparing results.

×